X-ray Si(Li) Detector with Electric Cooling ElSiX

Application

The X-ray SiLi detector with electric cooling is designed for X-ray fluorescence analyzers, X-ray diffractometers and X-ray spectrometers, where detector cooling by liquid nitrogen is undesirable or impractical.

Features

  • No liquid nitrogen
  • Excellent energy resolution
  • Thin Be windows
  • Allows for any spatial orientation of detection unit
  • Automatic restart after the blackout
  • Long-duration continuous functioning
  • Lightweight and compact design

COMPLETE SET

  • SiLi detector
  • Compressor unit
  • Connecting gas lines

DESCRIPTION

The X-ray Si(Li) detector with electric cooling is designed for X-ray fluorescence analyzers, X-ray diffractometers and X-ray spectrometers, where detector cooling by liquid nitrogen is undesirable or impractical.

SPECIFICATION

ParameterValue
2 Energy ab range, keV41 - 60
Detector sensitive
area, mm2
20 *
Energy resolution at 5.9 keV, eV
at shaping time constant 16 μs
at shaping time constant 1 μs
at count rate 105 cps and shaping time constant 1 μs

138
268
275
Peak / background ratio> 4000
Cooling time, h3
Ambient temperature, °С-5 to +38
Operating modeContinuous
Spatial orientation of detection unitAny
Voltage, V / frequency, Hz200 / 50-60
Power Supply, W< 60
Overall dimension, mm:
Detection unit
Compressor unit

80x135x150
140x160x300
Weight, kg
Detection unit
Compressor unit

1.8
7.5
Operating life time (MTTF), h85000
* Detectors with 20, 60 mm2 and 4 x 15 mm sensitive area are available by separate order.

FILES TO DOWNLOAD

PDF brochure