All-in-one Gamma Spectrometer with digital signal processing is applicable for HPGe detector spectrometric pulses registration, amplification, digital filtration and gamma-ray spectra registration. All-in-one Gamma Spectrometer is designed as one integrated monoblock being identical to standard laboratory detection unit with inbuilt preamplifier, however it has inbuilt MCA, HV supply, etc.
| Parameter | Value |
|---|---|
| ADC | 14bit, 10 MSps |
| Integral nonlinearity | <0.04% |
| Temperature stability | TK = 50 ppm |
| Digital signal processing | |
| Automated and manual adjustment of trigger threshold | |
| Fine gain adjustment (in steps of <0.05%) | 0.5 - 6.5 |
| Shaping time (Integration time / rise time of filter multiplied by 2) adjustable in the range (flat top adjustable 0 ... 15μs) | 0.1... 25.5μs |
| Channel splitting | 128, 256, 512, 1k, 2k, 4k, 8k, 16k |
| Differential nonlinearity (for 8k channels and 6μs shaping time) | <1% |
| Automated PZC adjustment, detector decay time constants from 40μs to 1ms can be compensated | |
| Double differential trigger filter | |
| Modes of operation - PHA (pulse height analysis), result is a histogram of pulse height. - MCS (multi channel scaling): registers number of counts in subsequent time intervals. - MCS mode allows to acquire an integral PHA spectrum in parallel to the MCS acquisition. - LLD / ULD settings are supported in both PHA and MCS mode. | |
| Energy resolution for HPGe detector with efficiency of 7%* at 122 keV at 1.33 MeV | 0,820 keV 1,76 keV |
| Throughput >100kcps in memory at 150kcps input rate and 0.2μs shaping time and 1.4μs flattop | |
| High Voltage Supply (programmable) | +/- 4500V, 0.1mA max |
| Power supply: USB, 4.5 V ... 5.25 V 280mA at full operation and high voltage 4500 V | |
| Computer Interface / Connector | USB |